Method Description

The determination of cannabinoid content in Cannabis and Cannabis Infused Products by High-Performance Liquid Chromatography with Diode Array Detection (HPLC-DAD) involves extracting cannabinoids from the sample and separating them through liquid chromatography. The diode array detector enables precise identification and quantification based on absorption spectra, ensuring accurate analysis of the product's cannabinoid profile for regulatory compliance and quality control in the cannabis industry.

Analyte List

Analyte Name LOD LOQ
CBC 0.437 ug/g 4.843 ug/g
CBD 0.643 ug/g 5.401 ug/g
CBDA 0.089 ug/g 0.295 ug/g
CBDV 0.092 ug/g 0.442 ug/g
CBG 0.207 ug/g 0.854 ug/g
CBGA 0.46 ug/g 4.815 ug/g
CBN 0.356 ug/g 4.316 ug/g
Delta-8 THC 0.404 ug/g 4.567 ug/g
Delta-8 THCV 0.05 ug/g 0.339 ug/g
Delta-9 THC 0.494 ug/g 1.516 ug/g
Delta-9 THCV 0.27 ug/g 0.403 ug/g
THCA 0.45 ug/g 4.947 ug/g
Total Cannabinoids N/A N/A
Total CBD N/A N/A
Total THC N/A N/A
Total THCV N/A N/A

Technical Data Sheet

Test Name - Common
Cannabinoid Potency Test
Method Name
Determination of Potency of Cannabis and Cannabis Infused Products ACN
Method Code
LAB-TM-025
Equipment
High-Performance Liquid Chromatography with Diode-Array Detection (HPLC-DAD)
Method Reference
Agilent Developed Method